Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

namic convenience kits labeled as: KIT Recalled by MEDLINE INDUSTRIES, LP - Northfield Due to The female luer fittings on the Pressure Monitoring...

Date: February 12, 2025
Company: MEDLINE INDUSTRIES, LP - Northfield
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact MEDLINE INDUSTRIES, LP - Northfield directly.

Affected Products

namic convenience kits labeled as: KIT,ARTERIAL,60,(152CM),PG; medical convenience kit, REF 70036160

Quantity: 40 units

Why Was This Recalled?

The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About MEDLINE INDUSTRIES, LP - Northfield

MEDLINE INDUSTRIES, LP - Northfield has 822 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report