Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medline Surgical Drapes: 1) SHEET Recalled by Medline Industries, LP Due to Medline has identified issues related to calibration of...

Date: January 7, 2026
Company: Medline Industries, LP
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medline Industries, LP directly.

Affected Products

Medline Surgical Drapes: 1) SHEET,DRAPE,40X58,STERILE, Model Number: DYNJP2410; 2) DRAPE,SHEET,ECONOMY,40X58",ST,20/CS, Model Number: DYNJP2411; 3) SHEET,DRAPE, 40X70,STERILE, Model Number: DYNJP2412; 4) DRAPE,MEDIUM,TAPE,40X70",ST,50/CS, Model Number: DYNJP2412T; 5) SHEET,DRAPE,53X77,STERILE, Model Number: DYNJP2414; 6) DRAPE,SHEET,3/4,53X77",ST,20/CS, Model Number: DYNJP2414R; 7) DRAPE,SHEET,ULTRAGARD,20/CS, Model Number: DYNJP2414UG; 8) DRAPE,RNF,53X77",ST,20/CS, Model Number: DYNJP2416; 9) SHEET,DRAPE,70X85,STERILE, Model Number: DYNJP2417; 10) DRAPE,FILM SHEET,44X65",ST,20/CS, Model Number: DYNJP2418; 11) DRAPE,SHEET,70X100",ST,12/CS, Model Number: DYNJP2419; 12) LEGGINGS,33X49",XL,PAIR,ST,CLR,20/CS, Model Number: DYNJP2462; 13) DRAPE,CHEST,FENESTRATION,15X10",ST,12/CS, Model Number: DYNJP2491; 14) DRAPE,MINOR PROC,6X6" FEN, ST,10/CS, Model Number: DYNJP2492; 15) SHEET, DRAPE, SPLIT, ST,10/CS, Model Number: DYNJP2498; 16) DRAPE,LAPAROTOMY,T-SHEET,ST,12/CS, Model Number: DYNJP3003; 17) DRAPE,LAPAROTOMY,T,ULTRAGARD,12/CS, Model Number: DYNJP3003UG; 18) DRAPE,LAPAROTOMY,POUCH,ST,12/CS, Model Number: DYNJP3008; 19) DRAPE,LAPAROTOMY,POUCH ULTRAGARD,12/CS, Model Number: DYNJP3008UG; 20) DRAPE,LAPAROTOMY,T,PEDIATRIC,ST,10/CS, Model Number: DYNJP3009; 21) DRAPE,LAPAROTOMY,ABDOMINAL,ST,12/CS, Model Number: DYNJP3101; 22) DRAPE,LAPROSCOPY,CHOLE,W/TROUGH,ST,8/CS, Model Number: DYNJP3102A; 23) DRAPE,LAP,CHOLE,W/TROUGH ULTRAGARD,12/CS, Model Number: DYNJP3102UG; 24) DRAPE,ABDOMINAL,MAJOR,ST,8/CS, Model Number: DYNJP3103; 25) DRAPE,ABDOMINAL,MAJOR,PCH/TRGH,ST,8/CS, Model Number: DYNJP3109; 26) DRAPE,TOP,102X53",ST,14/CS, Model Number: DYNJP4002; 27) DRAPE,CARDIOVASCULAR,SPLIT,ST,8/CS, Model Number: DYNJP4003; 28) DRAPE,ANGIO,BRACH,ECLIPSE,38X44,ST,40/CS, Model Number: DYNJP4106; 29) DRAPE,FEMORAL ANGIOGRAPHY,W/POUCH,8/CS, Model Number: DYNJP4119; 30) DRAPE,TOP,ARMCOVERS,106X59,STERILE,20/CS, Model Number: DYNJP4215; 31) DRAPE,UNDERBUTTOCK,PCH,ST,20/CS, Model Number: DYNJP6002; 32) DRAPE,UNDERBUTTOCK,GRAD POUCH,PORT,20/CS, Model Number: DYNJP6006; 33) DRAPE,EENT,SPLIT,ST,8/CS, Model Number: DYNJP7001; 34) SHEET,T,THYROID,ST,12/CS, Model Number: DYNJP7003; 35) DRAPE,EXTREMITY,89X128",ST,12/CS, Model Number: DYNJP8002; 36) DRAPE,EXTREMITY,ULTRAGARD,12/CS, Model Number: DYNJP8002UG; 37) T-DRAPE,EXTREMITY,ST,6/CS, Model Number: DYNJP8003; 38) T-DRAPE,EXTREMITY,ULTRAGARD,6/CS, Model Number: DYNJP8003UG; 39) DRAPE,HAND,ST,18/CS, Model Number: DYNJP8005; 40) DRAPE,HAND,ULTRAGARD,18/CS, Model Number: DYNJP8005UG; 41) DRAPE,HIP,W/POUCH,ST,5/CS, Model Number: DYNJP8201; 42) DBD-DRAPE,HIP,W/POUCHES, ULTRAGARD, Model Number: DYNJP8201UG; 43) DRAPE,ORTHOMAX,HIP,W/POUCH,6/CS, Model Number: DYNJP8211; 44) SHEET,ORTHO,SPLIT,ST,10/CS, Model Number: DYNJP8301; 45) DRAPE,SPLIT,77X120",10/CS, Model Number: DYNJP8304A; 46) ULTRAGARD SPLIT DRAPE,10/CS, Model Number: DYNJP8304UG; 47) DRAPE,SHOULDER,BEACH CHAIR,ST,5/CS, Model Number: DYNJP8412; 48) DRAPE,SHOULDER,BEACH CH,ULTRAGARD,5/CS, Model Number: DYNJP8412UG; 49) DRAPE,SHOULDER,BEACH CHAIR,ST,5/CS, Model Number: DYNJP8414; 50) DRAPE,LITHOTOMY,ST,8/CS, Model Number: DYNJP9001; 51) DRAPE,LAVH,ST,8/CS, Model Number: DYNJP9103; 52) DRAPE,VETERINARY,CUSTOM,SM,20/CS, Model Number: MED901; 53) DRAPE,VETERINARY,CUSTOM,M,20/CS, Model Number: MED902; 54) DRAPE,VETERINARY,CUSTOM,L,20/CS, Model Number: MED903

Quantity: 14,927,765 units

Why Was This Recalled?

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medline Industries, LP

Medline Industries, LP has 182 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report