Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medline convenience kits containing semi-rigid suction liners labeled as follows: Recalled by MEDLINE INDUSTRIES, LP - Northfield Due to Due to reported adverse events, Medline Industries, LP....

Date: June 27, 2024
Company: MEDLINE INDUSTRIES, LP - Northfield
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact MEDLINE INDUSTRIES, LP - Northfield directly.

Affected Products

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) MINOR BREAST, Pack Number DYNJ908104B; b) PLASTIC MAJOR, Pack Number DYNJ908106B; c) ABDOMINOPLASTY W/LIPO SUCTION, Pack Number DYNJ909946; d) PLASTIC, Pack Number DYNJ909951

Quantity: 350 units

Why Was This Recalled?

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About MEDLINE INDUSTRIES, LP - Northfield

MEDLINE INDUSTRIES, LP - Northfield has 822 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report