Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

CENTURION SnagFree Chest Tube insertion Tray Recalled by MEDLINE INDUSTRIES, LP - Northfield Due to One lot of chest tube insertion trays, containing...

Date: September 7, 2022
Company: MEDLINE INDUSTRIES, LP - Northfield
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact MEDLINE INDUSTRIES, LP - Northfield directly.

Affected Products

CENTURION SnagFree Chest Tube insertion Tray, Reorder CHT545

Quantity: 18 kits

Why Was This Recalled?

One lot of chest tube insertion trays, containing a silk suture component that is restricted to 1 cycle of sterilization, was processed though 2 cycles.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About MEDLINE INDUSTRIES, LP - Northfield

MEDLINE INDUSTRIES, LP - Northfield has 822 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report