Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ABX Pentra Calcium AS CP Recalled by MEDLINE INDUSTRIES, LP - Northfield Due to Product was shipped from a Medline warehouse to...

Date: October 22, 2021
Company: MEDLINE INDUSTRIES, LP - Northfield
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact MEDLINE INDUSTRIES, LP - Northfield directly.

Affected Products

ABX Pentra Calcium AS CP, Reagent 90 mL, REF A11A01954

Quantity: 6 boxes

Why Was This Recalled?

Product was shipped from a Medline warehouse to a customer at room temperature instead of refrigerated (ice). If affected product is used, the results from this impacted Calcium reagent will not be provided due to probable Quality Control failure. A delay in diagnosis may be experienced with repeat testing resulting in patient (customer) inconvenience, pain, and cost of the repeat blood draw and testing. All HII1220001954 product shipped/stored at a Medline branch between July 2021 and August 2021 in the identified lots were incorrectly stored and shipped. Six (6) boxes have been shipped to end users at room temperature instead of refrigerated (ice). Five (5) boxes are on hold at a Medline warehouse.

Where Was This Sold?

US Distribution to North Carolina.

About MEDLINE INDUSTRIES, LP - Northfield

MEDLINE INDUSTRIES, LP - Northfield has 822 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report