Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

NAMIC Manifolds. (1) NAMIC Angiographic MANIFOLD Recalled by Medline Industries Inc Due to Potential for sterile barrier breach. Microscopic pinholes within...

Date: August 6, 2021
Company: Medline Industries Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medline Industries Inc directly.

Affected Products

NAMIC Manifolds. (1) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 64038301, UPN H965640383011, 3 Valve Manifold, Off Star Handles, Ports on Right, Rx only, Sterile; (2) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 64038303, UPN H965640383031, 3 Valve Manifold, On Star Handles, Ports on Right, Rx only, Sterile; (3) NAMIC Angiographic MANIFOLD High Pressure, REF 70037200, UPN H965700372001, 2 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (4) NAMIC Angiographic MANIFOLD, High Pressure, REF 70037202, UPN H965700372021, 2 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile. (5) NAMIC Angiographic MANIFOLD, High Pressure, REF 70037301, UPN H965700373011, 3 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (6) NAMIC Angiographic MANIFOLD, High Pressure, REF 70037303, UPN H965700373031, 3 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (7) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038200, UPN H965700382001, 2 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (8) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038202, UPN H965700382021, 2 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (9) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038301, UPN H965700383011, 3 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; (10) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70038303, UPN H96570038301, 3 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (11) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70039202, UPN H965700392021, 2 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile; (12) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70039301, UPN H965700393011, 3 Valve Manifold, Off Handles, Ports on Right, Rx only, Sterile; and (13) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 70039303, 3 Valve Manifold, On Handles, Ports on Right, Rx only, Sterile.

Quantity: 816,420 manifolds and stopcocks

Why Was This Recalled?

Potential for sterile barrier breach. Microscopic pinholes within the sterile pouch could potentially lead to non-sterile conditions.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medline Industries Inc

Medline Industries Inc has 153 total recalls tracked by RecallDetector.