Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

TBS iNsight Version v.3.0.1 Product Usage: TBS iNsight is a Recalled by Medimaps Group Due to The FRAX adjusted for TBS values are not...

Date: October 3, 2016
Company: Medimaps Group
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medimaps Group directly.

Affected Products

TBS iNsight Version v.3.0.1 Product Usage: TBS iNsight is a medical device software that is installed on bone densitometers for analysis of bone microarchitecture and osteoporosis management.

Quantity: 15 units in US

Why Was This Recalled?

The FRAX adjusted for TBS values are not correct when: The FRAX feature is activated in TBS iNsight; and TBS has been computed from a spine scan where some vertebrae were excluded.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medimaps Group

Medimaps Group has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report