Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Rebstock Holding rod for fixation -For holding and fixation of Recalled by Medagent Inc Due to Holding arm could move (twist) uncontrollably in the...

Date: December 21, 2023
Company: Medagent Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medagent Inc directly.

Affected Products

Rebstock Holding rod for fixation -For holding and fixation of flexible Leyla Arms and spatula Art. No. 06-06-285 and 06-06-285MOD. surgical instrument.

Quantity: 84 units

Why Was This Recalled?

Holding arm could move (twist) uncontrollably in the front area. This could lead to a serious deterioration of patient's safety if the movement is transferred to the spatula, which is attached to the holding arm.

Where Was This Sold?

This product was distributed to 1 state: TX

Affected (1 state)Not affected

About Medagent Inc

Medagent Inc has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report