Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Abbott Point of Care Control Control i-STAT¿ Level 3 1.7 mL Recalled by Mckesson Medical-Surgical Inc. Corporate Office Due to Facility Critical Temperature Excursions on 08-JAN 2024 in...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Mckesson Medical-Surgical Inc. Corporate Office directly.
Affected Products
Abbott Point of Care Control Control i-STAT¿ Level 3 1.7 mL
Quantity: 2 units
Why Was This Recalled?
Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Mckesson Medical-Surgical Inc. Corporate Office
Mckesson Medical-Surgical Inc. Corporate Office has 276 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report