Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

McKesson Horizon Medical Imaging (HMI) products versions 4.6.1 to including Recalled by Mckesson Medical Immaging Due to McKesson has identified a design deficiency where under...

Date: February 18, 2016
Company: Mckesson Medical Immaging
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Mckesson Medical Immaging directly.

Affected Products

McKesson Horizon Medical Imaging (HMI) products versions 4.6.1 to including 11.9 and McKesson Radiology (MR) products 12.0 and 12.1.1.

Quantity: 631 devices

Why Was This Recalled?

McKesson has identified a design deficiency where under rare circumstances, imported images/studies may re-use a non-unique image directory. This issue may cause incorrect images to be displayed for a patient.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Mckesson Medical Immaging

Mckesson Medical Immaging has 4 total recalls tracked by RecallDetector.

Related Recalls