Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Mar Cor Purification Recalled by Mar Cor Purification Due to Mar Cor Purification manufactured Central Water Plant (CWP)...

Date: April 24, 2014
Company: Mar Cor Purification
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Mar Cor Purification directly.

Affected Products

Mar Cor Purification, Central Water Platform (CWP), 102, 104 and 106. Product Usage: The CWP is a Central reverse osmosis system with the intended use of producing water for dialysis of multiple patients.

Quantity: 197

Why Was This Recalled?

Mar Cor Purification manufactured Central Water Plant (CWP) reverse osmosis system, models 102, 104 and 106. This correction has been initiated due to a potential situation with a specific lot of the water inlet solenoid valve. The potential exists that the water inlet solenoid valve could cease to function. If this occurs, CWP will stop producing reverse osmosis water.

Where Was This Sold?

This product was distributed nationwide across all 50 states.