Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator Recalled by Maquet Cardiovascular Us Sales, Llc Due to The manufacturer has conducted a review of the...

Date: August 15, 2013
Company: Maquet Cardiovascular Us Sales, Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Maquet Cardiovascular Us Sales, Llc directly.

Affected Products

QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000. The QUADROX oxygenators are intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.

Quantity: 1,182 units

Why Was This Recalled?

The manufacturer has conducted a review of the results of product testing performed during the last twelve months and has identified specific lots of product which may not meet specifications for bond strength on the inlet and/or outlet connector.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Maquet Cardiovascular Us Sales, Llc

Maquet Cardiovascular Us Sales, Llc has 62 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report