Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433.01 Recalled by Maquet Cardiovascular Us Sales, Llc Due to MAQUET has identified a potential issue with the...

Date: April 26, 2013
Company: Maquet Cardiovascular Us Sales, Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Maquet Cardiovascular Us Sales, Llc directly.

Affected Products

MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433.01 EN 05. The YUNO OTN is used to support and position patients for surgical procedures immediately before, during and after the surgical phase. The table top is radio translucent and enables intraoperative use of X-ray equipment.

Quantity: 13 Tables; 24 Extension devices

Why Was This Recalled?

MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the Extension device issue: By fully actuating the lever of the extension device handle, the operator's skin can enter a gap between the handle lever and a recess in the handle bar. When the lever is released,

Where Was This Sold?

Worldwide Distribution - USA including Arizona, Georgia, North Dakota, South Carolina, Texas and Tennessee, Internationally to United Arab Emirates, Austria, Brazil, Canada, Switzerland, China, Germany, Spain, France, United Kingdom, Italy, Japan, Singapore, Thailand, Turkey, Serbia, and South Africa.

About Maquet Cardiovascular Us Sales, Llc