Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Getinge 88-Series-washer disinfectors intended for cleaning Recalled by Maquet Cardiovascular Us Sales, Llc Due to Potential leakage that may occur due to ruptured...

Date: February 10, 2020
Company: Maquet Cardiovascular Us Sales, Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Maquet Cardiovascular Us Sales, Llc directly.

Affected Products

Getinge 88-Series-washer disinfectors intended for cleaning, disinfection and drying Model Number: 88-5 Part Number: 88-103; 88-203

Quantity: 146 US

Why Was This Recalled?

Potential leakage that may occur due to ruptured spot welds in the chamber, could generate a situation for the user such as risk for slipping or infection from contaminated water

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Maquet Cardiovascular Us Sales, Llc

Maquet Cardiovascular Us Sales, Llc has 62 total recalls tracked by RecallDetector.

Related Recalls