Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

Getinge 46-Series Washer Disinfector Model # 46-4 Recalled by Maquet Cardiovascular Us Sales, Llc Due to Due to a missing water intake for Deionized...

Date: January 16, 2020
Company: Maquet Cardiovascular Us Sales, Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Maquet Cardiovascular Us Sales, Llc directly.

Affected Products

Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-502 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods.

Quantity: 4 units

Why Was This Recalled?

Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly installed. There have been no adverse events reported resulting in serious illness or injuries.

Where Was This Sold?

This product was distributed to 7 states: CA, CO, IA, MN, NY, SC, VA

Affected (7 states)Not affected

About Maquet Cardiovascular Us Sales, Llc

Maquet Cardiovascular Us Sales, Llc has 62 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report