Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Ultima OPCAB System Recalled by Maquet Cardiovascular, LLC Due to MAQUET has identified a potential issue with the...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Maquet Cardiovascular, LLC directly.
Affected Products
Ultima OPCAB System, Sterile, Rx Only, Product Usage: Intended for use during performance of minimally invasive cardiac surgery through a sternotomy incision approach. The ULTIMA Access RailPlatform in combination with the Activator II Drive Mechanism is used to spread the sternum, providing access and direct visualization to the thoracic cavity. The AccessRail Platform also allows for the organization of pericardial sutures. The ULTIMA Stablizer isolates and provides local immobilization of the target vessel on the beating heart.
Quantity: 20 units (Lot 25052176); 12 units (Lot #25052200)
Why Was This Recalled?
MAQUET has identified a potential issue with the ULTIMA OPCAB (part numbers OM-2003S & OM-2001D). Labeling for the 2 blade lots had been incorrectly applied. The standard blades from Lot #25052176 were incorrectly labeled as deep blades, and the 20 deep blades from Lot #25052200 were incorrectly labeled as standard blades.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Maquet Cardiovascular, LLC
Maquet Cardiovascular, LLC has 251 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report