Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for Recalled by Maquet Cardiovascular, LLC Due to Potential for there to be a fluid ingress...

Date: May 17, 2024
Company: Maquet Cardiovascular, LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Maquet Cardiovascular, LLC directly.

Affected Products

The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

Quantity: 37,186 units

Why Was This Recalled?

Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Maquet Cardiovascular, LLC

Maquet Cardiovascular, LLC has 251 total recalls tracked by RecallDetector.

Related Recalls