Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

CARDIOHELP Gas Bottle Holder HKH 8880 Recalled by Maquet Cardiopulmonary Ag Due to There is a possibility for the clamps of...

Date: April 4, 2018
Company: Maquet Cardiopulmonary Ag
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Maquet Cardiopulmonary Ag directly.

Affected Products

CARDIOHELP Gas Bottle Holder HKH 8880, Product Code/Part Number 70105.0901 Product Usage: The Cardiohelp Gas Bottle (or Cylinder) Holder HKH 8880 is an accessory for the CARDIOHELP System and required for inter-hospital patient transport in aircraft and road vehicles (mobile intensive care units and ambulances). Inter-hospital patient transport is transport of the patient between clinics, where the clinical environment is left (e.g., transport in vehicles such as ambulances or aircraft). The gas cylinder holder serves to hold an oxygen cylinder in aircraft.

Quantity: 14 units (1 US, 13 International)

Why Was This Recalled?

There is a possibility for the clamps of the gas bottle holder to crack or break when exposed to strong vibrations.

Where Was This Sold?

This product was distributed to 1 state: MO

Affected (1 state)Not affected

About Maquet Cardiopulmonary Ag

Maquet Cardiopulmonary Ag has 5 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report