Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ARIES SARS-CoV-2 Assay Recalled by Luminex Corporation Due to SARS-CoV-2 assay that detects SARS-CoV-2 ORF1ab and N...

Date: July 14, 2022
Company: Luminex Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Luminex Corporation directly.

Affected Products

ARIES SARS-CoV-2 Assay

Quantity: 257

Why Was This Recalled?

SARS-CoV-2 assay that detects SARS-CoV-2 ORF1ab and N genes may not detect the ORF1ab gene. A positive results if either SARS-CoV-2 genes is detected, but If both genes are mistakenly not detected, a false negative result would be generated.

Where Was This Sold?

This product was distributed nationwide across all 50 states.