Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The Livongo Health In Touch¿ Blood Glucose Monitoring System (BGMS) Recalled by Livongo Health Due to The instructional insert for the In Touch Control...

Date: February 18, 2016
Company: Livongo Health
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Livongo Health directly.

Affected Products

The Livongo Health In Touch¿ Blood Glucose Monitoring System (BGMS), is an Over-The-Counter (OTC) system designed for the self- monitoring of blood glucose by persons with diabetes in home settings to aid in their diabetes management.

Quantity: 63,000 units

Why Was This Recalled?

The instructional insert for the In Touch Control Solution for use with the In Touch meter and In Touch strips as a quality control check did not contain the limitation of sensor accuracy: (1) with the use of acetaminophen (Tylenol) (2) with the use of Ascorbic Acid (vitamin C) supplementation (3) in the presence of uric acid

Where Was This Sold?

US consignees: Arizona, California, Delaware, Florida, Georgia, Illinois, Kentucky, Maryland, Michigan, Minnesota, Mississippi, Missouri, Montana, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia and Washington.