Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
SORIN Heater-Cooler System 3T 120V~/60Hz REF 16-02-85 Description of product: Recalled by LivaNova USA Inc. Due to The firm has become aware that due to...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact LivaNova USA Inc. directly.
Affected Products
SORIN Heater-Cooler System 3T 120V~/60Hz REF 16-02-85 Description of product: The St¿ckert Heater-Cooler System is an independent (i.e. independent of the water supply) 3-circuit-heating/cooling system. The three separate water circuits have been developed to serve as: " Two circuits for the patient supplied by one tank (for the heating/cooling blanket and the heat-exchanger associated with the oxygenator), common temperatures ranging from 2 degrees Celsius to 41 degrees Celsius, tank volume 6 liters. " One interchangeable heating/cooling circuit for cardioplegia heat-exchanger, temperatures ranging from 2 degrees Celsius to 10 degrees Celsius (cooling tank) and/or from 15 degrees Celsius to 41 degrees Celsius (heating tank). If required, the circuits for the patient and the circuit for cardioplegia can be switched off separately, in order to increase the activated functional group s heating and/or cooling performance. Product Usage: The Stockert Heater-Cooler System 3T is intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.
Quantity: 8671
Why Was This Recalled?
The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. In addition, the firm is releasing a design upgrade to reduce the risk of potential emission of aerosols from the 3T.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About LivaNova USA Inc.
LivaNova USA Inc. has 28 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report