Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Liebel-Flarsheim Hydra Vision Urology X-Ray System (DR) Liebel-Flarsheim urology systems Recalled by LIEBEL-FLARSHEIM COMPANY LLC Due to Software issue. The difference between the display and...

Date: February 27, 2017
Company: LIEBEL-FLARSHEIM COMPANY LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact LIEBEL-FLARSHEIM COMPANY LLC directly.

Affected Products

Liebel-Flarsheim Hydra Vision Urology X-Ray System (DR) Liebel-Flarsheim urology systems facilitate radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation, primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/ surgery.

Quantity: 313

Why Was This Recalled?

Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.

Where Was This Sold?

This product was distributed to 30 states: AL, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IA, KS, KY, MD, MI, MN, NV, NH, NY, NC, OH, OK, PA, SC, TN, TX, VA, WA, WV, WI

Affected (30 states)Not affected

About LIEBEL-FLARSHEIM COMPANY LLC