Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.
GE Healthcare Pediatric Direct-Wired Disposable Defibrillation/Pacing Electrodes. REF 2059144-001 and Recalled by Leonhard Lang Gmbh Due to The firm discovered that during the use of...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Leonhard Lang Gmbh directly.
Affected Products
GE Healthcare Pediatric Direct-Wired Disposable Defibrillation/Pacing Electrodes. REF 2059144-001 and 2059144-005. . Product Usage: Usage: Skintact Pediatric Multifunction Electrodes are for use on pediatric patients, less than 8 years of age, or weighing less than 25 kg (55 lbs), for external defibrillation, pacing, monitoring and cardioversion. The device is non-sterile and for single use only.
Quantity: 1080 pieces
Why Was This Recalled?
The firm discovered that during the use of these defibrillation electrodes a possibility for arcing and a resulting malfunction exists. This could lead to a situation, in which a patient who is in a life threatening situation requiring a defibrillation can receive such therapy only with a delay or not at all.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Leonhard Lang Gmbh
Leonhard Lang Gmbh has 1 total recall tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report