Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
PROVEO 8 Ophthalmic Microscope System optical instrument for improving the Recalled by Leica Microsystems, Inc. Due to The observed issues can lead to unexpected interruptions...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Leica Microsystems, Inc. directly.
Affected Products
PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
Quantity: 142 units
Why Was This Recalled?
The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or sporadic failures in the proper on-screen display functionality of the HDR recording unit. Additionally, an anomaly in the device software may also lead to the unexpected behavior that when a user releases a handle or foot switch button this may not immediately end the actuated functionality. This behavior can only occur in unlikely workflows where a user changes a user profile or a mode change is initiated at the same time as a functional button is activated.
Where Was This Sold?
Worldwide
About Leica Microsystems, Inc.
Leica Microsystems, Inc. has 40 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report