Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Hydra Vision DR 60/80 System Recalled by Leibel-Flarsheim Company LLC Due to A complaint was received in which a hospital...

Date: April 25, 2013
Company: Leibel-Flarsheim Company LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Leibel-Flarsheim Company LLC directly.

Affected Products

Hydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used in radiologic imaging procedures.

Quantity: 12 units

Why Was This Recalled?

A complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum Fluoro Dose was found to be 30 R/min which is three times the Maximum Fluoro Dose limit of 10 R/min.

Where Was This Sold?

This product was distributed nationwide across all 50 states.