Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D Recalled by LAP of America Laser Applications, LLC Due to While projecting yellow lines it was notice that...

Date: March 21, 2014
Company: LAP of America Laser Applications, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact LAP of America Laser Applications, LLC directly.

Affected Products

LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view.

Quantity: 1 system

Why Was This Recalled?

While projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between the red and green projection was up to 2.5mm.

Where Was This Sold?

North Carolina

About LAP of America Laser Applications, LLC

LAP of America Laser Applications, LLC has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report