Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Pump Tube Set for MMS Libra and Solar Systems - Recalled by LABORIE MEDICAL TECHNOLOGIES, CANADA ULC Due to This field correction is being initiated to address...

Date: April 22, 2020
Company: LABORIE MEDICAL TECHNOLOGIES, CANADA ULC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact LABORIE MEDICAL TECHNOLOGIES, CANADA ULC directly.

Affected Products

Pump Tube Set for MMS Libra and Solar Systems - Product Usage: Intended for use during water perfusion urodynamic procedures to study the function of the bladder and urethra during filling and voiding, by monitoring pressure and pressure changes.

Quantity: 480 units

Why Was This Recalled?

This field correction is being initiated to address potentially damaged seals in the packaging of some the Laborie Pump Tube Sets. Finished goods inventory was detected with damage to their seals in the form of breaches or folds. Based on our investigation the issue appears to be limited to a single lot of part number 902101226 pump tubing sets.

Where Was This Sold?

This product was distributed nationwide across all 50 states.