Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Torque Limiting Adaptor Recalled by Laboratoires Bodycad, Inc. Due to There is a risk of breakage of the...

Date: July 2, 2021
Company: Laboratoires Bodycad, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Laboratoires Bodycad, Inc. directly.

Affected Products

Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.

Quantity: 10 in the U.S.

Why Was This Recalled?

There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.

Where Was This Sold?

This product was distributed to 3 states: CO, MI, NY

Affected (3 states)Not affected

About Laboratoires Bodycad, Inc.

Laboratoires Bodycad, Inc. has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report