Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Stage-1 Tissue Punch Recalled by Keystone Dental Inc Due to The cutting edge of tissue punches may be...

Date: January 24, 2018
Company: Keystone Dental Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Keystone Dental Inc directly.

Affected Products

Stage-1 Tissue Punch, Disposable, 4.5mm; catalog number: ISS141K. Intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations.

Quantity: 101

Why Was This Recalled?

The cutting edge of tissue punches may be potentially damaged with small dents and nicks.

Where Was This Sold?

Worldwide Distribution -- US, to the state of California; and, the countries of Canada and Germany

About Keystone Dental Inc

Keystone Dental Inc has 16 total recalls tracked by RecallDetector.