Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Genesis Surgical Cassette Tapered Implants are intended for placement following Recalled by Keystone Dental Inc Due to The surgical cassette contained extra-large white grommets making...

Date: March 13, 2018
Company: Keystone Dental Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Keystone Dental Inc directly.

Affected Products

Genesis Surgical Cassette Tapered Implants are intended for placement following natural tooth loss or for immediate placement into an extraction socket

Quantity: 19

Why Was This Recalled?

The surgical cassette contained extra-large white grommets making the 3.8 mm and 6.5mm dental burrs wobble within the grommets in the surgical cassette.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Keystone Dental Inc

Keystone Dental Inc has 16 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report