Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Image1 SPIES X-LINK Camera Control Unit (CCU) Recalled by Karl Storz Endoscopy Due to Potential electrical interference issue when the Image1 SPIES...

Date: June 20, 2016
Company: Karl Storz Endoscopy
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Karl Storz Endoscopy directly.

Affected Products

Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera head, Product Model No. TH116 & TH115, and DC1 Video Mediastinoscope

Quantity: 619 units total (90 units in US)

Why Was This Recalled?

Potential electrical interference issue when the Image1 SPIES X-LINK and Image1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinocope and an electro-surgery generator (ESG).

Where Was This Sold?

This product was distributed to 21 states: AR, CO, CT, FL, GA, ID, IL, KY, MD, MA, MI, MN, MO, NJ, NM, NY, PA, TN, TX, WI, DC

Affected (21 states)Not affected

About Karl Storz Endoscopy

Karl Storz Endoscopy has 70 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report