Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Justright Surgical Recalled by Justright Surgical, LLC Due to JustRight Surgical is recalling specific lots of JustRight...

Date: August 27, 2014
Company: Justright Surgical, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Justright Surgical, LLC directly.

Affected Products

Justright Surgical, JustRight 5mm Stapler, REF JR-ST25-2.0, Sterile, EO, Rx Only, JR-ST25-2.0 product is packaged in cases of 6. The JustRight 5mm Stapler is intended for use in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomosis.

Quantity: 3 cases containing 18 units

Why Was This Recalled?

JustRight Surgical is recalling specific lots of JustRight 5mm Stapler and Reload products because they may deliver poor distal staple formation, past the end of the knife cut.

Where Was This Sold?

Distributed in the states of Colorado, California and New York.

About Justright Surgical, LLC

Justright Surgical, LLC has 4 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report