Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
TECNIS Eyhance IOLs are permanent intraocular implants intended for implantation Recalled by Johnson & Johnson Surgical Vision Inc Due to Nonconforming product was distributed in error. Product was...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Johnson & Johnson Surgical Vision Inc directly.
Affected Products
TECNIS Eyhance IOLs are permanent intraocular implants intended for implantation in the capsular bag. The TECNIS Simplicity Delivery System is designed as a single-use disposable preloaded modular cartridge that functions as both the primary packaging and as a sterile, disposable insertion system for delivering the IOL into the eye during cataract surgery.
Quantity: 1 lens
Why Was This Recalled?
Nonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal endotoxin specification limit and the FDA recommended endotoxin specification limit. Product was released in error based on the results of additional testing performed during the investigation into release testing nonconformance.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Johnson & Johnson Surgical Vision Inc
Johnson & Johnson Surgical Vision Inc has 7 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report