Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MyWay Walker Chassis - Product Usage: is intended for use Recalled by James Leckey Design Ltd Due to Due to a failure of the welds, the...

Date: March 1, 2021
Company: James Leckey Design Ltd
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact James Leckey Design Ltd directly.

Affected Products

MyWay Walker Chassis - Product Usage: is intended for use by children already using a supportive walking device.

Quantity: 2,719 walkers

Why Was This Recalled?

Due to a failure of the welds, the casters detached from the rear legs of the walker.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About James Leckey Design Ltd

James Leckey Design Ltd has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report