Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part) Recalled by Isopure Corp Due to There is a potential for an uneven plastic...

Date: April 19, 2019
Company: Isopure Corp
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Isopure Corp directly.

Affected Products

Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part)

Quantity: 8

Why Was This Recalled?

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Where Was This Sold?

This product was distributed to 13 states: AL, FL, GA, MD, MA, MN, NM, NY, NC, OH, OK, TX, VA

Affected (13 states)Not affected

About Isopure Corp

Isopure Corp has 5 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report