Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

IBA Proton Therapy System - PROTEUS 235. PTS-11 versions Recalled by Ion Beam Applications S.A. Due to The patient positioning system can perform motions in...

Date: September 18, 2023
Company: Ion Beam Applications S.A.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ion Beam Applications S.A. directly.

Affected Products

IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient Positioning System.

Quantity: 5 (US), 9 (OUS)

Why Was This Recalled?

The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.