Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Xper Flex Cardio Physiomonitoring System Recalled by Invivo Corporation Due to The Invasive Blood Pressure (IBP) numeric values will...

Date: September 24, 2018
Company: Invivo Corporation
Status: Completed
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Invivo Corporation directly.

Affected Products

Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388

Quantity: 76 units

Why Was This Recalled?

The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard

Where Was This Sold?

Distributed to the Netherlands

About Invivo Corporation

Invivo Corporation has 19 total recalls tracked by RecallDetector.