Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Invivo Sentinelle Venguard Breast MRI Auxiliary Table/Tabletop with 8/16 Channel Recalled by Invivo Corporation Due to A diode was inadvertently not assembled into the...

Date: May 6, 2016
Company: Invivo Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Invivo Corporation directly.

Affected Products

Invivo Sentinelle Venguard Breast MRI Auxiliary Table/Tabletop with 8/16 Channel Coil Array for GE 1.5T/3T MRI Systems.

Quantity: 3

Why Was This Recalled?

A diode was inadvertently not assembled into the system, which may result in coil overheating

Where Was This Sold?

This product was distributed to 1 state: WI