Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Invacare¿ Pronto¿ Air Personal Transporter (power wheelchair) Recalled by Invacare Corporation Due to The batteries used in the Invacare¿ Pronto¿ Air...

Date: February 2, 2015
Company: Invacare Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Invacare Corporation directly.

Affected Products

Invacare¿ Pronto¿ Air Personal Transporter (power wheelchair), Invacare Corporation, One Invacare Way, Elyria, Ohio, 44035-4190, (800) 333 -6900.

Quantity: 1,057 finished devices

Why Was This Recalled?

The batteries used in the Invacare¿ Pronto¿ Air Personal Transporter models power wheelchairs have the possibility to overheat during charging which could potentially result in leaking battery acid and/or swelling battery, loss of use, chemical burns and/or property damage.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Invacare Corporation

Invacare Corporation has 37 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report