Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with Recalled by Invacare Corporation Due to Power Wheelchairs with a LiNX Gyro component running...

Date: November 12, 2021
Company: Invacare Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Invacare Corporation directly.

Affected Products

Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R, SRX-20SP, SRX-20MP

Quantity: 1 unit US

Why Was This Recalled?

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Invacare Corporation

Invacare Corporation has 37 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report