Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Component- Dianz Capacitor used in PerfectO2 Oxygen Concentrator Recalled by Invacare Corporation Due to The firm has found that the Component- Dianz...

Date: October 31, 2014
Company: Invacare Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Invacare Corporation directly.

Affected Products

Component- Dianz Capacitor used in PerfectO2 Oxygen Concentrator, packaged one unit per box.

Quantity: 138,446 units distributed globally

Why Was This Recalled?

The firm has found that the Component- Dianz Capacitor has ruptured which results in a potential fire hazard.

Where Was This Sold?

The product was distributed throughout the U.S. and to foreign countries, including Canada and other countries TBD. The affected product was also distributed to eight government consignees.

About Invacare Corporation

Invacare Corporation has 37 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report