Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Aquatec Ocean VIP Recalled by Invacare Corporation Due to The snap-on fixtures of the backrest cushion on...

Date: August 31, 2013
Company: Invacare Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Invacare Corporation directly.

Affected Products

Aquatec Ocean VIP, Aquatec Ocean Dual-VIP, Aquatec Ocean E-VIP mobile shower and toilet commode with tilting function. Mobile shower and toilet commode with tilting function

Quantity: 538 (US only) 3317 (Total)

Why Was This Recalled?

The snap-on fixtures of the backrest cushion on the Mobile Shower and Toilet Commode Aquatec Ocean VIP and Ocean Dual VIP do not withstand a required pull out force of 500 N, when the product is used in a reclined position and or tilted position. This may result in the snap-on fixture coming unlatched.

Where Was This Sold?

This product was distributed to 1 state: PA

Affected (1 state)Not affected

About Invacare Corporation

Invacare Corporation has 37 total recalls tracked by RecallDetector.