Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Aquatec Fixed Backrest - accessory for Mobile Shower and Toilet Recalled by Invacare Corporation Due to The Aquatec¿ Fixed Backrest withstands a load of...

Date: June 25, 2014
Company: Invacare Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Invacare Corporation directly.

Affected Products

Aquatec Fixed Backrest - accessory for Mobile Shower and Toilet commode Aquatec Ocean, Ocean VIP and Ocean E-VIP family.

Quantity: 465 units

Why Was This Recalled?

The Aquatec¿ Fixed Backrest withstands a load of only 250 newtons at uneven load. This may result in the Fixed Backrest becoming detached from the frame.

Where Was This Sold?

This product was distributed to 1 state: PA

Affected (1 state)Not affected

About Invacare Corporation

Invacare Corporation has 37 total recalls tracked by RecallDetector.

Related Recalls