Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
da Vinci Xi Surgical System Surgical System Recalled by Intuitive Surgical, Inc. Due to The firm become aware of a potentially defective...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Intuitive Surgical, Inc. directly.
Affected Products
da Vinci Xi Surgical System Surgical System, IS4000, 00886874114216 - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.
Quantity: 123 surgical systems
Why Was This Recalled?
The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Intuitive Surgical, Inc.
Intuitive Surgical, Inc. has 244 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report