Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

da Vinci SP Surgical systems with the following Model Name/ Recalled by Intuitive Surgical, Inc. Due to Issue was identified during internal engineering evaluation. Use...

Date: September 22, 2021
Company: Intuitive Surgical, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Intuitive Surgical, Inc. directly.

Affected Products

da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605 ASSY,SSC,SP1098,P4/ 380940-44 / 00886874114735 ASSY,VSS,SP1098,P4/ 380941-44 / 00886874114742 SOFTWARE, SYSTEM EMBEDDED RLS, SP1098, B70_P4_B327 (OS4 v4.0.0 (P4_B327) / 610112-327 / N/A

Quantity: 46

Why Was This Recalled?

Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may require surgeon to resolve an injury or need moderate intraoperative intervention. Also, use may potentially cause user frustration or a minor delay to troubleshoot.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Intuitive Surgical, Inc.

Intuitive Surgical, Inc. has 244 total recalls tracked by RecallDetector.