Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

8MM Recalled by Intuitive Surgical, Inc. Due to Due to an increase in complaints for pitch...

Date: December 19, 2024
Company: Intuitive Surgical, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Intuitive Surgical, Inc. directly.

Affected Products

8MM, Small Graptor REF 470318 Is a grasping retractor Instrument used with the da Vinci X/Xi Surgical Systems in a wide variety of procedures to perform a variety of surgical tasks including manipulation and retraction of tissue.

Quantity: 42028 devices

Why Was This Recalled?

Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Intuitive Surgical, Inc.

Intuitive Surgical, Inc. has 244 total recalls tracked by RecallDetector.