Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

G3 Opticage Expandable Interbody Fusion Device Recalled by Interventional Spine Inc Due to During implantation, the proximal end of the device...

Date: November 22, 2016
Company: Interventional Spine Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Interventional Spine Inc directly.

Affected Products

G3 Opticage Expandable Interbody Fusion Device, Model No. 9180-00, 9180-01, 9180-06, 9180-07, 9180-17, 9101-00, 9101-06 The Opticage Expandable Interbody Fusion Device is a posterior lumbar intervertebral body fusion device and is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Opticage Interbody Fusion Device can be implanted via posterior, transforaminal or lateral approach. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Candidates for surgery should be skeletally mature and have had six months of conservative treatment. These patients may have had primary or secondary surgery, but no previous fusion at the involved levels. The device is not intended to be used as a stand-alone device. It must be used with supplemental internal spinal fixation systems that have been cleared for use in the lumbar spine (i.e. facet screw fixation systems, facet compression devices and posterior pedicle screw and rod systems).

Quantity: 239 units

Why Was This Recalled?

During implantation, the proximal end of the device could separate if the angle between the device and insertion instrument is too steep. This separation will prevent the device from expanding.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Interventional Spine Inc