Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Integra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal Recalled by Integra LifeSciences Corporation Due to As a result of the investigation into an...

Date: May 9, 2013
Company: Integra LifeSciences Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Integra LifeSciences Corporation directly.

Affected Products

Integra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia.

Quantity: 692

Why Was This Recalled?

As a result of the investigation into an adverse trend of complaints for reports of misalignment of the calcaneal or tibial screws during use of the Panta Nail Support Device, Integra has decided to issue this voluntary medical device correction. The misalignment can result in the inability to properly insert the screws into the Panta Nail and create the potential need for revision surgery.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Integra LifeSciences Corporation

Integra LifeSciences Corporation has 29 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report