Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MEDIHONEY Dressing with Active Leptospermum Honey Wound and Burn Dressing Recalled by Integra LifeSciences Corp. Due to There is a potential for pinholes in the...

Date: September 13, 2024
Company: Integra LifeSciences Corp.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Integra LifeSciences Corp. directly.

Affected Products

MEDIHONEY Dressing with Active Leptospermum Honey Wound and Burn Dressing W/APPLICATOR, 1.5oz tube (Model Number 31515) and 3.5oz tube (Model Number 31535)

Quantity: 120124 units

Why Was This Recalled?

There is a potential for pinholes in the applicator pouch film which creates a sterility concern.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Integra LifeSciences Corp.

Integra LifeSciences Corp. has 216 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report