Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Recalled by Integra LifeSciences Corp. Due to Due to a complaint, it was determined that...

Date: November 7, 2014
Company: Integra LifeSciences Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Integra LifeSciences Corp. directly.

Affected Products

Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX¿ Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX¿ Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX¿ Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX¿ Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse;schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247 <eilkendorf Germany Distributed by Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536, USA

Quantity: 2,506 units

Why Was This Recalled?

Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.

Where Was This Sold?

This product was distributed nationwide across all 50 states.